BFM-700M Made in China Fetal Maternal Monitor LCD LED Monitor with Twin

BFM-700M Made in China Fetal Maternal Monitor LCD LED Monitor with Twin

Model No.︰BFM-700M

Brand Name︰Bestman

Country of Origin︰China

Unit Price︰US $ 2055 / pc

Minimum Order︰1 pc

Inquire Now

Product Description

Feature:

5.7 inches TFT LCD displays 2 x US and 1x UC data, mother's NIBP, TEMP, RR, SpO2, and ECG main parameters and monitoring trace curve etc.;

 

FHR collects signal indication, alarm indication for abnormal FHR, fetal movement indication, and can freeze monitoring status;

 

Can set the time limit (0-99 minutes), automatically entering dormant state if no signal input;

 

Store monitoring data for 24 hours, with replay function;

 

Coding keyboard adjusts all parameters setting of the unit, convenient to use;

 

Adopt 832 thermal linear array printer, store the data and print rapidly; and printing width is 112mm;

 

The printer parameters can be adjusted according to user needs, including printing paper speed, trace line thickness, signal sampling rate, upper and lower alarming limit, real time, and date;

 

Set-up data and parameters can be automatically saved after power off;

 

The monitoring data can be conveyed to exterior computer through RS-232 or USB port;

 

Trolly (adjustable) for option, convenient to move and operate.

 


Specifications:
1.  US
Transducer   9 crystals, wide wave beam
Working mode: Pulsed ultrasound Doppler
Ultrasonic frequency: 1.0MHz±10%
Measuring range:    50-240 BPM
Alarm range:  Up: 55-240 BPM  Adjustable continuously within range
                       Low: 50-235 BPM  Adjustable continuously within range

 

2.  TOCO
Display range: 0-100units  (non-linear error: ≤±10%)

 

3.  PC PORT
Edition USB 1.1 or USB 2.0; RS-232

 

4. THERMAL RECORDER
Paper width : 112 ± 2mm effective record width : 105 ± 2mm;  Paper length : 15m
Recorder density :  US:  27 BPM/ CM ;   UC:  40 BPM / CM
Printing speed (adjustable):  1 cm/min, 2 cm/min, 3 cm/min

 

5. POWER and BATTERY
Battery: 16.8V Ni-MH rechargeable batteries group;  AC 220 /110 V ;  Consume power: ≤ 60 W
Continuous working time in static condition: ≥ 1 hours

 

6.ECG and HR
Measuring range: 30 BPM -350 BPM
Cardiac electric noise level: ≤ 30 μV
Cardiac electric frequency response:
Filtering mode: 1-25 MHz
Non-filtering mode: 0.05-100 MHz
Cardiac electric Input loop current: ≤ 0.1 μA

 

7.NIBP
Measuring range:
a) SBP: (8.034.0) kPa or (60255) mmHg
b) DBP: (4.026.0) kPa or (30195) mmHg
c) MAP: (5.328.7) kPa or (40215) mmHg
Automatic measuring interval: 3 min-99 min adjustable continuously within range
interval error: ≤ ±10 s

 

8.RR
Measuring  range: 0 BPM-100 BPM

 

9.TEMP
Channels: 1
TEMP probe: surface probe 1pcs
Measuring range: 28-45
Response time: ≤30s

 

10.SpO2
Measuring range: 0-99%

 

11.PR
Measuring range: 30-250 BPM

 

12.ENVIROANMENT REQUIREMENT
Working Temperature: +5 -40   Humidity: ≤80%  Atmospheric pressure: 86 kPa106 kPa

 

13.ACCESSORIES

2 x US probe, 1 pcs TOCO probe, 1 pcs RS-232 serial port line, 1 pcs probe bandage, 1 pcs SpO2  probe, 1 pcs TEMP probe, 1 pcs electrode probe, 1 pcs power cable, 1 pcs power switch, 1 pcs BP cuff, event marker, 2 pcs printing paper, 250ml gel, , warranty card, operation manual, service manual

Product Image